dBadge2 / dBadge2 IS

INTERESTED?   CONTACT US


Product Overview 

Transmit sound level measurements through Bluetooth® 4.0 wireless connectivity to increase productivity—no need to stop working to download readings! Dose, status, level alerts and more can be accessed through the free Airwave app and sent directly to your Apple or Android enabled device. Casella's dBadge2 noise dosimeters capture every possible noise exposure related parameter in a single measurement run, and make it available for subsequent download and analysis. Email summary data, photographs, and even notes for review or inclusion in reports.

Ensure the correct hearing protection is chosen for each application and location. The measurements taken are in accordance with CFR 1910.95, ISO 9612:2009, and UK Guidance document L108. The LCD displays current noise levels in color-coded format for easy interpretation of alarm conditions at a glance. Use the preset alarm limits to adhere to OSHA, ISO, MSHA, and ACGIH guidelines, or set your own.

Each personal noise dosimeter comes with two mounting clips, one at the top and one at the bottom, for secure shoulder attachment and features a reliable and robust 1/2" diameter microphone with windscreen. Order a kit to get started and add additional units as needed to cover an entire fleet!

dBadge2 Personal Noise Dosimeter Kits allow remote start, stop, or pause of a measurement run without interrupting the user—ideal for hazardous locations, positions, and filed data capture.

This item is intended as an add on to a kit. This single meter does not include a docking station/calibrator, charger, USB cable for direct data transfer to PC, Insight data analysis and summary software, a case to house up to 10 dosimeters, or access to the free Airwave app for your handheld device. Please order a kit to get started.


Technical specifications 

Standards: ANSI S1.25:1991 R2007, IEC 61252 Ed 1.1 (2002-3)

Linear Operating Range: 55.0-140.3 dB (A) RMS 

Peak Measurement Range: 90.0-143.3dB (C or Z weighted)

Sound Exposure Range: 0.0- 6,100.0 Pa2Hours 

Frequency Weightings: A, C and Z, Type 2 

Time Weightings: Fast, Slow and Impulse 

Exchange Rate: Q=3 or Q=5dB exchange rates 

Threshold and Criterion: 70-90dB in 1dB steps 

Operating Temperature Range: 0°C to +40°C (for <±0.5dB error limit) -10°C to +50°C (for <±0.8dB error limit) 

Ambient Pressure Range: <±0.5dB over 85 - 112kPa 

Humidity Range: <± 0.5dB over 30% - 90% (non-condensing) 

Storage Temperature Range: -10°C to +50°C, 

Battery: Internal NiMH, 800mAH 

Run Time: Typically 35 hours using Broadband Mode Typically 15 hours using Octave mode 

Charge Time: <6 Hours from fully discharged

Maximum Run Duration: 24 hours 

Maximum number of Runs: 100 

Storage Capacity: Internal USD Flash memory - 300 hours of run data (including 1 second profile data) and up to 90 minutes of event audio recording

PC Interface: USB 2.0 Full speed 12Mbps via Docking Station 

Dimensions: 85L x 54W x 55H mm 

Weight: 117g 

Environmental I.P. Rating: IP65 (with permeable air vent)